    Jul 12, 2018


    When testing the vacuum system of a mixed gas with a heat conduction vacuum gauge, if the volume percentage Vi (i = 1, 2, 3, ..., n) of various gases in the mixed gas is known, the relative sensitivity S of mixed gas for the air can be calculated by the following formula.

     

    blob.png 


    Where: Vi—the volume percentage of the i-th gas;


    Sri—the relative sensitivity of the i-th gas.

     

    blob.png


    After calculating Sr, the result measured by the above method is multiplied by Sr to obtain the actual vacuum of the mixed gas.