Boost test - Determine if there is a real leak in the vacuum system
Mar 27, 2025| Boost test - Determine if there is a real leak in the vacuum system
The pressure rise test is also called the pressure retention test or the vacuum retention test. When the valve between the vacuum pump and the chamber is closed under vacuum, the pressure of the chamber will rise (rebound) due to venting, virtual leakage or leakage. By dividing the pressure increase by the elapsed time, the pressure rise rate of the vacuum system can be calculated and plotted as the curve shown in the following figure. Pressure rising and falling speed is usually expressed as a Pa/hr, for general industrial vacuum equipment, pressure test, pressure rise rate is more than a Pa/hr needs to be investigated and solve the problem, some of the high vacuum equipment, need 0.5 Pa/hr or less pressure rise rate. As shown above in system A, if measuring the pressure rise rate constant many times, it is likely a system with real leaks. Correspondingly, for system B in the figure above, as time goes by, its pressure rise rate decreases, that is, the increase rate of system pressure slows down with time, suggesting that there is virtual leakage or venting (there may also be a small leak). Pressure drop and pressure increase tests won't locate leaks, and will only indicate all gas sources (the actual leak and deflated or virtual leak), the stacking effect. If a real leak is suspected, the next step is usually to detect the leak using a helium leak detector.

IKS PVD company, decorative coating machine, tools coating machine, DLC coating machine, optical coating machine, PVD vacuum coating line, and a turn-key project are available. Contact us now, Email: iks.pvd@foxmail.com